American Weigh Scales TS-500 User Manual

Page 4

background image

Place the required weight on the scale.

5. Press ON/TARE once more. If the display shows

AL F, calibration is complete! AL E means calibration

failure - Repeat steps 1 - 5.

B. TS-100 Calibration

TS-100 - uses a 100g calibration weight

1. Turn OFF the scale.

2. Press and hold the [MODE] key and [POWER/

TARE] key at the same time until the display shows a

random number.

3. Release the [POWER/TARE] key and [MODE] keys.

4. Press the [POWER/TARE] key again. The display

will show AL 0 followed by the required calibration

weight. Place the required weight on the scale.

5. Once the scale is stable, the LCD will display AL F.

Calibration is complete! AL E means calibration failure

(Repeat steps 1 - 5)

VI. TROUBLESHOOTING

The primary reasons for inaccuracy or malfunction

are low batteries, incorrect calibration, overload or

operating on an unstable surface.

[Display] Elimination of errors

[LO-]

Replace the batteries.

[O-Ld]

You are overloading the scale. Remove the

excessive load immediately!

[AL E]

A calibration error has occured.