Taring weight display (TARE)

The TARE function is useful if an additional weight should not be considered while weighing.

Proceed as follows:
– Press the Start key with no load on the

scale.

– First, place the additional weight onto

the scales and activate the TARE func-
tion by pressing down and briefly hold-
ing down the

HOLD/TARE button.

The display is reset to zero.

0.0

appears

in the display. The note “NET” appears.

– Now place the patient onto the scales

with the previously weighed additional

weight.
The scales show the weight of the pa-
tient without the additional weight. You

can now perform as many weighing

operations as you like.

– Deactivate the TARE function by press-

ing the button down and holding it

briefly once again; this puts the scales

back into the normal weighing mode,

the tare value (the additional weight val-
ue) is deleted.

– The tare value is likewise erased when

the scales are switched off.

Store weight value (HOLD)

You can also continue displaying the weight determined once the load has been removed.

This makes it possible to look after a patient first, before having to note the weight:
– Press the

HOLD/TARE button while

the scales have a weight on it.

– “HOLD” appears in the digital display.

The weight remains stored in the dis-
play until the Hold function is automati-
cally switched off.

– The triangle with the exclamation mark

indicates that this frozen display is an

additional facility which is not subject to

calibration.

– The HOLD function can be switched off

by pressing the

HOLD/TARE key

again.

0.0

appears in the display if the

scales have no load, otherwise the cur-
rent weight is shown. The “HOLD” dis-
play vanishes.

Note: The HOLD function can also be used if the TARE function has been execu-
ted beforehand.