(f) Turn-off timer
• The instrument will automatically turn off if it has not received

laser scanning signal for 5 minutes

(g) Illuminator system
• There is an illuminator system at the right-faced display window.

Press the illuminator button which will illuminate the display window.

8. Self-Check and Calibration

X-Direction Accuracy Self-Check

1. For clarity, we define the direction of handle as X-direction,

and another direction as Y-direction

2. Place the unit on a platform that is 30 feet (10m) away from a

wall indoors, with the handle facing the wall head-on. Unlock
the unit and adjust it to the quick-scan status.

3. Draw one vertical line on exactly front wall, which intersection

with the laser line can be marked as point A.

(Note: This test should be done indoors with dim lighting It’s
critical that the laser mark is easily seen)

4. Turn the instrument by 180 degrees, mark the intersection of

vertical line and laser line as point B, and also mark the center
of point A and point B as point 0.

Wall

40-6520_6530 4/20/05 4:12 PM Page 15