1.2 Technical

Specifications

Table 1. CBCT Technical Specifications

TYPE Programmable DC Power Supply

PHYSICAL SPECIFICATIONS 19”W x 7”H x 15”D, (48 cm x 17 cm x 38 cm); Weight: 25 lbs (11.3 kg)

INPUT VOLTAGE

20 – 300 Vdc, 20A fuse

OUTPUT VOLTAGE 5% – 95% of input voltage with 2% regulation at max current (up to 80A)

OUTPUT VOLTAGE CHANNELS One continuous and one pulse DC output

OUTPUT VOLTAGE

PROTECTION

Shutdown if current exceeds 80A or if current drawing time exceeds 500
ms @ 80A

VOLT METER RANGE Input/output meters, 0 – 300 Vdc; Accuracy: 1% of reading, ±0.2V

TIMER DISPLAY RANGE 0.0000 – 999.999 seconds; Accuracy: 0.1 ms, ±1 digit

TIMER START INPUT Internal (coil initiation), wet-contact input (10 c 240 Vac/dc), or dry-contact

input (1 – 200 ohms)

TIMER STOP INPUT Wet-contact (10 – 240 Vac/dc) or dry-contact (1 – 200 ohms)

DISPLAY Back-lit LCD screen (20 characters by 4 lines); viewable in both bright

sunlight and low-light levels

SAFETY Designed to meet UL 6101A-1 and CAN/CSA C22.2 No 1010.1-92

standards

ENVIRONMENT Operating: -10°C to 50°C (+15°F to +122°F); Storage: -30°C to70°C (-22°F

to +158°F)

CABLES Three 6-foot (#10 AWG) DC cable sets, Two 10-foot timer cable sets, One

ground cable, one cable bag

OPTIONS Transportation case

WARRANTY One year on parts and labor