Field calibration, Co2 field calibration procedure – TSI IAQ50 User Manual
Page 29

29
especially important in applications where 
strict calibration records must be main-
tained. 
Send the instrument to TSI Instruments 
Ltd. prepaid. Securely package your 
instrument in a strong container sur-
rounded by at least 5 cm of suitable shock-
absorbing material. Include a purchase 
order that clearly shows the instrument 
model number and serial number, a 
contact name, phone, fax number, and 
RMA number. Mark the outside of your 
shipping container with the RMA number. 
This will expedite processing of your 
instrument when we receive it. 
 
Field Calibration 
CO
2
Field Calibration Procedure
1. Turn the IAQ50 on and set the Baro-
metric Pressure to the current Baro-
metric Pressure. Be sure to use “station 
pressure” which may be available at 
the local airport. “Station pressure” is 
not corrected to sea level. 
2. Turn the IAQ50 off and flip DIP switch
#7 to ON.