Low flow applications (5 to 500 ml/min), Setup - low flow – SKC 224-52 AirChek 52 User Manual
Page 11
9
Low Flow Applications (5 to 500 ml/min)
Using a CPC and Single Adjustable Low Flow Holder
Figure 1
Front, back, and top views of AirChek 52 Sampler
For additional drawings, see pages 19 and 21.
Setup - Low Flow
Ensure that the battery is fully
charged before sampling. See page
5 for details.
After charging the
battery pack, it is good
practice to run the pump for
approximately fi ve minutes
before calibrating. This ensures
the battery is in more steady-
state conditions and improves
the agreement in pre and post-
sampling calibrations.
1
Charger and battery
connected
1 LCD
2 Flow adjustment screw
3 On/Off
button
4 Pump intake/inlet hose
barb
5 Protective control cover
6 Belt
clip
7 Charging
jack
8 Accessory
mounting
screws
1
3
4
5
6
7
8