Error codes – Robinair 16900 Refrigerant Identifier User Manual
Page 17

16900 Refrigerant Identifier
15
DETECTOR FAULT: Signifies internal infrared sensor failure due to
temperature extremes.
Corrective Actions:
1. Allow the instrument to stabilize at room temperature for at least
30 minutes before attempting use.
2. To clear the error code, power down the instrument. The error code
automatically clears.
CALIBRATION FAULT: Signifies an infrared calibration fault.
Corrective Actions:
1. Verify the air intake port and sample exhaust port of the instrument
are not clogged or obstructed.
2. If the instrument is being used in an enclosed area, refrigerant
vapors may build up in the surrounding atmosphere and be drawn
into the intake port during calibration. Move the instrument to a
location with fresh ambient air, free of refrigerant vapors.
Use the instrument in locations that provide adequate ventilation to
prevent the accumulation of refrigerant vapors in the surrounding
atmosphere.
3. Hang the unit so there is a flow of fresh air around it.
4. To clear the error code, power down the instrument. The error code
automatically clears.
Error Codes