MAESTRO User’s Manual Revision 8.1
8
Accuracy
3
1.0 %
50 µV < 6 kHz
2%
50 µV 6kHz to 10 kHz
Default Trigger Level
2 %
SOFTWARE
TRIGGER LEVEL
0.1% to 99.9%, 0.1% resolution
Repetition
Frequency
4
2 kHz for data acquisition in the real time full screen display, no missing
point
1.3 kHz for data acquisition in real time with time stamp, no missing point
2 kHz in statistics mode, no missing point
From 2 kHz to 10kHz the MAESTRO will send the latest energy pulse
every 500us. (corresponding to a subsampling at 2 kHz).
Frequency
Measurement
Accurate frequency measurement up to 14kHz
0-1000 Hz: 0.1% accuracy
1000-2000 Hz: 0.5%
2000-14000 Hz: 1%
Statistics
Current value, Max, Min, Average, Std Dev., RMS stability, PTP stability,
Time, Pulse #, Repetition Rate, Avg Power.
Data Storage
Continuous on USB stick
General Specifications
Digital Display
112.9 x 84.7mm RGB color LCD with touchscreen, 640 x 480 Pixels
Display Rate
3 Hz numeric display
15 Hz bar graph & needle display
Scope
500 divisions
Data Displays
Real time, Scope, Averaging, Statistics, Digital tuning needle
User input correction
factors
1 multipliers and 1 offsets (7 digits floating point)
Analog Output
0
– 1 volt, full scale, ± .5%
Internet Upgrades
USB STICK
PC Serial Commands
USB & Ethernet
5
& RS-232
No missing point
throughput serial
300 Hz
3
Including linearity, detector dependant.
4
In a full size statistic’s windows, no scope display and no serial communication.
5
USB cable included. RS-232 cable must be purchased separately.
6
Backlight at 50% in real time display. The scope and averaging displays decrease the autonomy by
20%.