beautypg.com

Appendix 1: data output specifications – Fairbanks Ultegra Health Scale User Manual

Page 8

background image

Appendix 1:

Data Output Specifications

This appendix contains detailed data output information on the Ultegra.

RS232-C Serial Data Output

A.

Introduction

The single RS232-C output port is designed for interface with customer’s data
processing equipment.

B.

Specifications

• RS232-C compatible Data Signal
• 7 data bits
• Odd Parity
9600 Baud Rate (+0.1%)
• Two Stop Bits
• US-ASCII Character Set
• Mark = -3 to -15V
• Space = + 3 to +15V
• Maximum distance of 50 cable-feet
• Connector (at end of the attached cable)

DB9 Female for connection to a PC serial port

C.

Character Frame

Characters are transmitted in an ASCII format at 9600 baud (+ 0.1%). The receiver
must be capable of a tolerance of 9600 (+/- 2.5%) baud to allow for line losses and
frequency skew. Character frame consists of one start bit, 7-bit character length, odd
parity bit and two stop bits.

08/08

8

50735 Rev. 6

3 to 15 VDC

ON State (Spacing)

Logic State

-3 to -15 VDC

OFF State (Marking)

S

T

A

R

T

B

I

T

0

ON

OFF OFF

D

0

L

S
B

D

1

D

2

D

3

D

4

D

5

D

6

P
A

R

I

T

Y

B

I

T

S

T

O

P

B

I

T

0

S

T

O

P

B

I

T

0