Pump Removal/Installation

(TURBO

2

P only)

Removal

1. Loosen the cover plate mounting screws and remove the instrument

coverplate.

2. Disconnect the pump motor wires from the circuit terminals.

3. Tag the pump inlet and outlet tubings and then disconnect both from

the pump assembly.

4. Loosen and remove the four screws securing the pump assembly to

the instrument chassis and remove the pump.

Installation

1. With the instrument cover plate removed, set the pump assembly in

place.

2. Insert and tighten the four screws to secure the pump assembly to

the instrument chassis.

3. Connect the tagged pump inlet and outlet tubings to the pump

assembly.

4. Connect the pump motor wires to the circuit terminals.

5. Replace the instrument cover plate and then insert and tighten the

cover plate mounting screws.