Series 4060

Operation

Teledyne

Analytical

Instruments

49

2. The MAIN MENU and SUB MENU screens have a flashing

arrow on the left side indicating which item is being pointed
to.

3. MODIFIABLE

ITEM screens have a flashing value

somewhere in the line.

4.6.4.1

M

AIN

M

ENUS

The MAIN MENU consists of 15 functions you can use to customize

and check the operation of the analyzer. They are listed here with brief
descriptions:

SPAN:

Set up and/or start a span calibration

ALT-SPAN

: Set up and/or start an alternate span calibration.

MODEL:

Displays Manufacturer, Model, and Software

version of the instrument.

SELF-TEST:

The instrument performs a self-diagnostic

routine to check the integrity of the power supply, output
boards, cell and amplifiers.

ALARMS:

Used to set the alarm setpoints and determine

whether each alarm will be active or defeated, HI or LO
acting, and failsafe or not.

RANGE:

Used to set up three analysis ranges that can be

switched manually.

 ANALOG-OUT ADJUST: Adjust the analog output with

offset and scaling.

LINEARIZATION

: Linearization option.

CHANGE STREAM

: (optional feature) Controls a relay

switch for using alternate gases. Has no effect during span
mode (span gas relay setting used at that time).

AUTO-CAL:

Used to define and/or start an automatic

calibration sequence. AUTOCAL is an optional feature not
included in the standard configurations of the 4060.

TIMING

: Low level application specific timing parameters.

Factory set. Do not change these.