Teledyne 3000MA - Paramagnetic oxygen analyzer Addendum User Manual
Page 2

Addendum for Model 3000MA Percent Oxygen Analyzer
Page 2
Teledyne Analytical Systems
Software
The software for this instrument adds hold timers for span and zero
calibration; each calibration has its own timer separate timer.
Span Delay:
By pressing the Span key, the operator accesses the setup mode of
the span calibration. After selecting AUTO span calibration the
following screen will be the hold timer adjustment:
Span Delay: 5 min.
The hold time for the span can be adjusted between 1 minute and 60
minutes. Five minutes is the factory default. When changed the value
is stored so it can be recovered after power down.
Zero Delay:
By pressing the Zero key, the operator accesses the setup mode of
the zero calibration. After selecting AUTO zero calibration the
following screen will be the hold timer adjustment:
Zero Delay: 5 min.
The hold time for the span can be adjusted between 1 minute and 60
minutes. Five minutes is the factory default. When changed the value
is stored so it can be recovered after power down.
Reference Drawings
D74505
Outline Diagram
B74504
Piping Diagram
Spare Parts
A78648
EPROM Software Assembly
Project Engineer ________________ Date ______________