PRODUCT SPECIFICATION DATA



Ranges:

0-10, 0-100, 0-1000, 0-10000 ppm Oxygen

CAL range for air calibration


Sensitivity: 0.5% of Full Scale


Accuracy:

±2% of full scale (except ±1 ppm for 0-10 ppm range) at constant

temperature and pressure (temperature and pressure of calibration).


±5% of full scale (except ±1 ppm for 0-10 ppm range), over

operating temperature range (once temperature equilibrium has been

achieved).

System Operating Temperature: 0-50

° C


Response Time:

90% response time 61 seconds

Reproducibility:

±1% at constant temperature


Sensor Type: Class

B-2C


System Power Requirements:

Model 311 - AC power for battery recharge circuit of two

current limited rechargeable NiCad batteries, 115 VAC,

50/60Hz

(100/220

VAC

optional).


Weight:

6 lb (2.71 Kg.).


Approval:

Intrinsically Safe (Class I, DIV. I, GROUP A, B, C and D), Factory

Mutual

approved.