Since no moving parts are used in the analyzer, no routine mainte-

nance is required, other than checking the gas flow rate periodically.

Symptom

Symptom

Symptom

Symptom

Symptom

Cause

Cause

Cause

Cause

Cause

What

What

What

What

What T

T

T

T

To Do

o Do

o Do

o Do

o Do

a) Bad fuse.

b) The power supply of +15V

and -15V may be defective.

c)

The bridge power supply
(100 mA) may be defective.

a) Blocked vent may be

causing back-pressure in
the measurement cell,
resulting in high cell pres-
sure, erratic readings, and a
constricted flow.

a) The temperature controller

may not be working.

1. No output.

2. Output drifting very slowly.

3. Output goes high slowly when
the instrument is first turned on.

a) Check and replace fuse.

b) If faulty, change the power

supply board.

c)

Measure it, and if faulty,
change the voltage regula-
tor/amplifier board.

a) Clean the vent.

a) Check that the bulb above

the power switch is flashing,
and if the bulb is not flash-
ing, change the temperature
controller board.

b) If the temperature controller

is working, perform the
same checks as in #2
above.