I: “new jersey” thermocouple option, Ersey, Hermocouple – Perma Pure eCOOL 5000 Series User Manual
Page 18: Ption

Sect
ion I: “New Jersey” Thermocouple Option
18
I:
“N
EW
J
ERSEY
”
T
HERMOCOUPLE
O
PTION
Some air quality management districts (e.g., those in New Jersey and Southern
California) require temperature measurement of the gas stream at the outlet of the
last heat exchanger on the cooler. Perma Pure offers a 1/32-inch diameter
hypodermic-style type K thermocouple that can be inserted into a special heat
exchanger (i.e., it has a small port for insertion of the thermocouple) so the actual
sample dew point temperature can be measured. This is sometimes referred to as
the New Jersey thermocouple outlet temperature option.
If the cooler is run as a single stream, the final heat exchanger will have a New
Jersey thermocouple to sense the temperature inside the heat exchanger. A
quantity of two (2) NJ thermocouples are required when the cooler is used for dual-
stream simultaneous conditioning.
In the Baldwin
™-Series eCOOL 5000™, this New Jersey Thermocouple can be
connected to one of the extra thermocouple inputs for readout. Alternately, for a 4-
20mA output, the optional isolator board should be purchased.
eCOOL™ 5000 Series New Jersey Thermocouple Option
Part No.
Description
3KTC-001
Kit, Thermocouple, NJ Option, Type K, w/Fitting & Sleeve
3KTC-002
Kit, Thermocouple, NJ Option, Type J, w/Fitting & Sleeve
3CCB-026
Circuit Board, eCooler, Isolator Board