Section E: Start-up Procedure

8

E:

S

TART

-

UP

P

ROCEDURE

Plug in the power cord to a properly grounded main circuit. The Ready Green LED
will come on within 3 minutes, indicating the relay temperature (10°C) has been
achieved. After approximately 3 minutes, the set point of +5°C. (41°F) is achieved.
The SLIP Green LED should also be on if a water slip sensor is connected to the
internal relay board. The sample gas flow may be started immediately after the
READY Green LED comes on.

The Model M325 is virtually maintenance free. However, in the event of electrical
problems, refer to the troubleshooting guide in this manual. All voltages can be read
at the PCB terminal strip. Any deviations from the correct voltages indicate a
problem.