beautypg.com

2 internal, direct insertion – Orion Instruments Jupiter 2xx FOUNDATION Fieldbus User Manual

Page 10

background image

10

57-649 Jupiter Magnetostrictive Transmitter - F

OUNDATION

fieldbus™

2. Position the Jupiter transmitter on the side of the MLI

where it will be attached. Mark the location and the exact
area where the clamps will be attached to hold the Jupiter
in place.

3. Attach the lower clamp and tighten so that it remains in

place, but loose enough so that there is still room to place
the guide tab from the Jupiter between the inside of the
clamp and the outer diameter of the MLI chamber. See
Figure 1.

4. The upper clamp will need to be open to a large enough

diameter to be able to mount to the MLI as well as the
probe. The upper clamp should be positioned just above
the

3

4

" NPT threads. See Figure 2.

5. Mount the Jupiter guide pin in the lower clamp and tight-

en. If necessary, use strapping tape to temporarily hold in
place on the MLI. See Figure 1.

6. Position the upper clamp to attach the unit to the MLI

and tighten. See Figure 1.

7. Discard any tape temporarily holding the Jupiter to the

MLI.

2.4.2 Internal, Direct Insertion

Use caution when handling probes to ensure probe is not
bent during installation. A bend in the probe may prevent
float from traveling freely up and down the probe.

1. Verify float will pass through vessel opening, if not, it will

be necessary to attach the float after the probe is installed.

2. Carefully insert probe into vessel and thread or bolt to the

mating connection as appropriate.

3. The float is held on the probe by a C-clip inserted into a

groove machined into the tip of the probe. The float is
attached or removed by removing and reinserting the
C-clip. See Figure 3. To ensure proper float orientation,
the float is marked “Up ”.

Bottom view

Up

Figure 3

Float Attachment Detail

Figure 2

Upper Clamp

©