Nova-Tech GKa Indicator User Manual
Page 53

51 © Adam Equipment Company 2011
Apply power to the indicator. If the jumper has been used the display will ask for a
code number, “Pn “ on the Weight Display immediately. Or press the [Tare] key
during the time the calibration counters are being displayed.
Enter the number 2006 when “Pn “ is displayed and then press [Tare].
The displays will show the first parameters, called “F1” “CAL”.
To select another parameter press the [Func] key to advance through the
parameters.
Press the [Tare] key to enter a parameter.
To exit a parameter, press the [Zero] key.
The display will show the parameter number and a name.
When a parameter is entered by pressing the [Tare] key, the displays will guide you
through the parameter selected and the options available.
The parameters available are:
“F1 CAL”
To enter the Calibration
“F2 dEC”
Decimal Point Position
“F3 CAP”
Default Weighing Unit
“F4 Int”
Initial Zero Range
“F5 rEZ”
Re-Zero range
“F6 SCS”
Successive Tare Enable
“F7 Cnt”
Display the A/D counts
“F8 Zem”
Zero Mode
“F9 Lvd”
Low voltage detection
“F10 Cn”
Calibration and Parameter counters