7.1.2 MEA1060-Up
Operating temperature
10 °C to 50 °C
Storage temperature
0 °C to 50 °C
Dimensions (W x D x H)
165 mm x 165 mm x 14.5 mm (19 mm with sockets)
Weight
600 g
Supply voltage (external power supply)
± 6 VDC to ± 9 VDC
Supply current
max. 220 mA, typically ± 150 mA
Number of stimulus input channels
60
Maximum recommended stimulus input
voltage
± 5 V
Number of analog input channels
60
Input voltage
-4 mV to 4 mV (with respect to the standard gain of
1200)
Input impedance
10
11
parallel to 10 pF
Input noise
< 800 nV
RMS
Noise density
e
n
=
Number of analog output channels
60
Output voltage
± 5 V
Output current
10 mA
Output impedance
300
Bandwidth
1 Hz to 3000 Hz (standard), 1 Hz to 5000 Hz (custom)
Gain
1200 (standard), 500 to 5000 (custom)
Microscope compatibility
Upright microscopes
Distance focal plane to microscope table
8 mm
7.1.3 Heating
Element
The heating element is mounted onto the bottom side of the base plate of the MEA1060
amplifier. It is used for keeping the temperature in the MEA culture chamber stable during
recordings.
Input voltage
max 10 V
Input current
max 2 A
Temperature sensor
PT 100
Resistance
15 ± 2
Heating temperature
Ambient temperature (min 10 °C) to 50 °C
Accuracy
0.1 °C (in the range of 20 °C to 40 °C)
Recovery time
0.5 min to 2 min
Calibration constant
T = c * (R R
0
) / R
0
for c = 259.7 °C
Thermal resistance
6 °C / Watt
Typical time constant
50 s