ADPT-NN-64
Adapter for connecting two MPA32Is to the
Acute 64 Probe from NeuroNexus Technology
Warning: The ADPT-NN-64 may only be used together with the MPA32I from Multi Channel
Systems MCS GmbH, and only for the specified purpose. Damage of the device and even
injuries can result from improper use.
!
Top Side
PCB Top View (Probe on the bottom side)
°
°
1
2
1
2
°
°
2
1
1
2
Top Side
MPA32I
No. 1
MPA32I
No. 2
ADPT-NN-64
Acute 64 Probe (NeuroNexus Technology)
Adapter Top Side
Adapter Bottom Side
Direction
of the ----->
Electrodes
Schematic diagram :
Connect the male pins at the bottom side of the Acute 64 Probe to the female pins at the bottom side
of the adapter ADPT-NN-64.
Connect the upper output pins at the top side of the adapter ADPT-NN-64 to the MPA32I No. 1.
It is mandatory, that the top side of the MPA32I No. 1 faces upwards. (The side with no screws
is considered the top side of the MPA32I).
Connect the lower output pins at the bottom side of the adapter ADPT-NN-64 to the MPA32I No. 2.
The top side of the MPA32I No. 2 must face upwards to the bottom side of the MPA32I No.1,
otherwise the preamplifiers will not work.
Acute 64
Probe
ADPT-NN-64
MPA32I No. 1
MPA32I No. 2
Acute 64 Probe Electrodes
ADPT-NN-64