beautypg.com

M4 series specifications – Meriam M4 Series Precision Calibrator Data Logger Quick Use Guide User Manual

Page 2

background image

Meriam Process Technologies • 10920 Madison Avenue • Cleveland, Ohio 44102 U.S.A.

Phone (216) 281-1100 • Fax (216) 281-0228 • www.meriam.com

24V Source: Typical Load Line

17.0

18.0

19.0

20.0

21.0

22.0

23.0

24.0

25.0

0

5

10

15

20

25

Current (mA)

O

u

tp

u

t V

o

lt

s

(V

)

M4 Series SPECIFICATIONS


Pressure Measurement

M400: single pressure sensor (specified by customer)
M402: dual pressure sensors – any two compound gauge or absolute sensors, mix and match as desired
Accuracy: ±0.025% of Reading from 10 - 100% FS, ±0.002% of Full Scale under 10% of FS
Temperature Performance: Accuracy includes all affects of temperature from -20º to +50º C (-4º to +122º F)


Pressure Limits

DN sensors:

2x range when pressurized on P1 (HI) side only, 150 PSI

when applied simultaneously to P1 (HI) and P2 (LO) sides.

DI sensors:

3x range when pressurized on P1 (HI) side only,
3x range or 150 PSI (whichever is less) on P2 (LO) side only

1000 PSI when applied simultaneously to P1 (HI) and P2 (LO) sides.

CI, AI sensors: 2x range

Media Compatibility

DN sensors:

Non-isolated for clean, dry, non-corrosive gases only (Brass, 316L SS, Silicon gel)

DI sensors:

Isolated for fluids compatible with 316L SS and Viton

CI, AI sensors: Isolated for fluids compatible with 316L SS

Voltage & Current, Measurement, Sourcing, Sinking

DC Voltage: ±50 VDC measure, 0 – 24 VDC source
DC Current: ±100 mA DC measure, 0 – 24 mA DC source, 1 – 24 mA sink at 1-50 VDC external power
Accuracy: ± (0.015% of Reading +0.002 units)
Temperature Performance: Accuracy includes all affects of temperature from -20º to +50º C (-4º to +122º F)

Resolution (measure mode): 0.001 V, 0.001 mA or 0.0001 V, 0.0001 mA (user selectable)


Loop Power

250Ω HART® resistor included
See 24 VDC Load Curve at right
Four (4) AA alkaline batteries power a transmitter at
20 mA output for 10 hours


Temperature

Operating: -4 to 122°F (-20 to 50°C)

Storage: -40 to 140°F (-40 to 60°C)

Connections

Pressure:

1

8

” NPT (female)

Electrical: Standard banana jacks on ¾” centers
Firmware updates, data retrieval: USB mini-B (female)

Power

Portable service: Four (4) AA alkaline batteries provides 30+ hours of continuous measurement service
Operation with PC: Power over USB using USB type A to mini-B cable (requires high power 500 mA port)

Enclosure: 8.5” L x 3.75” W (max.) x 2.25” D (max.), polycarbonate case, Softflex

®

over-molded bumpers, IP40

Weight: M400 single sensor: 1.5 – 1.8 lbs, M402 dual sensor: 1.8 lbs

Consult manual located on the product CD or at

www.meriam.com

for detailed

instructions, software utility, and safety information!