M4 series specifications – Meriam M4 Series Precision Calibrator Data Logger Quick Use Guide User Manual
Page 2
Meriam Process Technologies • 10920 Madison Avenue • Cleveland, Ohio 44102 U.S.A.
Phone (216) 281-1100 • Fax (216) 281-0228 • www.meriam.com
24V Source: Typical Load Line
17.0
18.0
19.0
20.0
21.0
22.0
23.0
24.0
25.0
0
5
10
15
20
25
Current (mA)
O
u
tp
u
t V
o
lt
s
(V
)
M4 Series SPECIFICATIONS
Pressure Measurement
M400: single pressure sensor (specified by customer)
M402: dual pressure sensors – any two compound gauge or absolute sensors, mix and match as desired
Accuracy: ±0.025% of Reading from 10 - 100% FS, ±0.002% of Full Scale under 10% of FS
Temperature Performance: Accuracy includes all affects of temperature from -20º to +50º C (-4º to +122º F)
Pressure Limits
DN sensors:
2x range when pressurized on P1 (HI) side only, 150 PSI
when applied simultaneously to P1 (HI) and P2 (LO) sides.
DI sensors:
3x range when pressurized on P1 (HI) side only,
3x range or 150 PSI (whichever is less) on P2 (LO) side only
1000 PSI when applied simultaneously to P1 (HI) and P2 (LO) sides.
CI, AI sensors: 2x range
Media Compatibility
DN sensors:
Non-isolated for clean, dry, non-corrosive gases only (Brass, 316L SS, Silicon gel)
DI sensors:
Isolated for fluids compatible with 316L SS and Viton
CI, AI sensors: Isolated for fluids compatible with 316L SS
Voltage & Current, Measurement, Sourcing, Sinking
DC Voltage: ±50 VDC measure, 0 – 24 VDC source
DC Current: ±100 mA DC measure, 0 – 24 mA DC source, 1 – 24 mA sink at 1-50 VDC external power
Accuracy: ± (0.015% of Reading +0.002 units)
Temperature Performance: Accuracy includes all affects of temperature from -20º to +50º C (-4º to +122º F)
Resolution (measure mode): 0.001 V, 0.001 mA or 0.0001 V, 0.0001 mA (user selectable)
Loop Power
250Ω HART® resistor included
See 24 VDC Load Curve at right
Four (4) AA alkaline batteries power a transmitter at
20 mA output for 10 hours
Temperature
Operating: -4 to 122°F (-20 to 50°C)
Storage: -40 to 140°F (-40 to 60°C)
Connections
Pressure:
1
⁄
8
” NPT (female)
Electrical: Standard banana jacks on ¾” centers
Firmware updates, data retrieval: USB mini-B (female)
Power
Portable service: Four (4) AA alkaline batteries provides 30+ hours of continuous measurement service
Operation with PC: Power over USB using USB type A to mini-B cable (requires high power 500 mA port)
Enclosure: 8.5” L x 3.75” W (max.) x 2.25” D (max.), polycarbonate case, Softflex
®
over-molded bumpers, IP40
Weight: M400 single sensor: 1.5 – 1.8 lbs, M402 dual sensor: 1.8 lbs
Consult manual located on the product CD or at
www.meriam.com
for detailed
instructions, software utility, and safety information!