Discontinued – Max Machinery 276-5XX SERIES 4-20MA TRANSMITTERS User Manual
Page 3

Specifications:
Range (Flowmeter RPM)
For a 4-20mA Output
276-515 and 276-525 . . . . . . . . . . . . . . . . . . . . . . . 21 RPM Minimum
3600 RPM Maximum
276-517 and 276-527 . . . . . . . . . . . . . . . . . . . . . . . .18 RPM Minimum
3600 RPM Maximum
Accuracy
Linearity (Input RPM vs Output Voltage)
Typical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .± 0.05% or ± 0.01mA
(whichever is larger)
Maximum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .± 0.1% or ± 0.01mA
(whichever is larger)
Zero Stability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .± 0.01mA
Temperature Range (See Temperature Considerations)
Electronics:
Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -10°C to 65°C
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -40°C to 70°C
Stator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -40°C to 130°C
Temperature Coefficients
Flowmeter Magnet . . . . . . . . . . . . . . . . . . . . . . . . . ± 0.5% per 100°C
Transmitter
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ± 0.2% per 100°C
Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . + 0.01mA per 100°C
Supply Voltage
Maximum at 25mA . . . . . . . . . . . . . . . . . . . . . . . . 35V (Continuously)
50V (5 Sec or less)
Minimum at 4mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.5 Volts
at 20mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.5 Volts
276-515-350 © 2006 Max Machinery, Inc.
( 3 )
SPECIFICATIONS
Discontinued