Laurel Electronics Laureate Rate Meter & Totalizer with Functions A+B, A-B, AxB, A_B, A_B-1 User Manual
Page 2
LAUREL
ELECTRONICS INC., 3183-G Airway Ave., Costa Mesa, CA 92626, USA • Tel 714-434-6131 • www.laurels.com 2
Specifications
Display
Readout
Display Range
Zero Adjust
Span Adjust
Indicators
6 LED digits, 7-segment, 14.2 mm (.56"), red or green LED
-999999 to +999999, XXXXEX notation beyond 999999
-999999 to +999999
0 to 999999
Four LED lamps
Inputs
Types
Signal Ground
Channel A Frequency
Channel B Frequency
Minimum Signal
Maximum Signal
Noise Filter
Contact Debounce
AC, pulses from NPN, PNP transistors, contact closures, magnetic pickups
Common ground for channels A & B
0.005 Hz to 1 MHz
0.005 Hz to 250 kHz
Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V)
250 Vac
1 MHz, 30 kHz, 250 Hz (selectable)
0, 3, 50 ms (selectable)
Rate Accuracy
Time Base
Span Tempco
Long-term Drift
Crystal calibrated to ±2 ppm
±1 ppm/°C (typ)
5 ppm/year
Power
Voltage, standard
Voltage, optional
Frequency
Power Isolation
85-264 Vac or 90-300 Vdc (DC operation not UL approved)
12-32 Vac or 10-48 Vdc
DC or 47-63 Hz
250V rms working, 2.3 kV rms per 1 min test
Excitation Output (standard)
5 Vdc
10 Vdc
24 Vdc
Output Isolation
5 Vdc ± 5%, 100 mA
10 Vdc ± 5%, 120 mA
24 Vdc ± 5%, 50 mA
50 Vdc to meter ground
Analog Output (optional)
Output Levels
Current compliance
Voltage compliance
Scaling
Resolution
Isolation
4-20 mA, 0-20 mA, 0-10V, -10 to +10V (single-output option)
4-20 mA, 0-20 mA, 0-10V (dual-output option)
2 mA at 10V ( > 5 kΩ load)
12V at 20 mA ( < 600Ω load)
Zero and full scale adjustable from -99999 to +99999
16 bits (0.0015% of full scale)
250V rms working, 2.3 kV rms per 1 min test
(dual analog outputs share the same ground)
Relay Outputs (optional)
Relay Types
Current Ratings
Output common
Isolation
2 Form C contact relays or 4 Form A contact relays (NO)
2 or 4 Form A, AC/DC solid state relays (NO)
8A at 250 Vac or 24 Vdc for contact relays
120 mA at 140 Vac or 180 Vdc for solid state relays
Isolated commons for dual relays or each pair of quad relays
250V rms working, 2.3 kV rms per 1 min test
Serial Data I/O (optional)
Board Selections
Protocols
Data Rates
Digital Addresses
Isolation
Ethernet, Ethernet-to-RS485 converter, USB, USB-to-RS485 converter, RS485
(dual RJ11), RS485 Modbus (dual RJ45), RS232.
Modbus RTU, Modbus ASCII, Laurel ASCII protocol
300 to 19200 baud
247 (Modbus), 31 (Laurel ASCII),
250V rms working, 2.3 kV rms per 1 min test