Specifications

Display

Readout
Range
Indicators

6 LED digits, 7-segment, 14.2 mm (.56"), red or green
-999999 to +999999
Four LED lamps

Inputs

Types
Grounding
Minimum Signal
Maximum Signal
Noise Filter
Contact Debounce

AC, pulses from NPN, PNP transistors, contact closures, magnetic pickups.
Common ground for channels A & B
Nine ranges from (-12 to +12 mV) to (+1.25 to +2.1V).
250 Vac
1 MHz, 30 kHz, 250 Hz (selectable)
0, 3, 50 ms (selectable)

Stopwatch Mode

Timing Modes:
With Ch A only
With Ch A tied to Ch B
With Ch A and Ch B

Timing Interval
Timing Resolution
Selectable Decimal Time
Selectable Clock Time
Output & Display Update


+ to + edge, or - to - edge.
+ to - edge, or - to + edge.
+ edge of A to + edge of B, + edge of A to - edge of B,
- edge to A to - edge of B, - edge of A to - edge of B
1 µs to 999,999 hrs
0.2 µs to 1 hr
999999 H, M or S format with decimal point
HH.MM.SS format
30 ms + gate time programmable from 10 ms to 199.99 s

Accuracy

Time Base
Span Tempco
Long-term Drift

Crystal calibrated to ±2 ppm
±1 ppm/°C (typ)
±5 ppm/year

Power

Voltage, standard
Voltage, optional
Frequency
Power Isolation

85-264 Vac or 90-300 Vdc (DC operation not UL approved)
12-32 Vac or 10-48 Vdc
DC or 47-63 Hz
250V rms working, 2.3 kV rms per 1 min test

Excitation Output (standard)

5 Vdc
10 Vdc
24 Vdc
Output Isolation

5 Vdc ± 5%, 100 mA
10 Vdc ± 5%, 120 mA
24 Vdc ± 5%, 50 mA
50 Vdc to meter ground

Analog Output (optional)

Output Levels

Current compliance
Voltage compliance
Scaling
Resolution
Isolation

4-20 mA, 0-20 mA, 0-10V, -10 to +10V (single-output option)
4-20 mA, 0-20 mA, 0-10V (dual-output option)
2 mA at 10V ( > 5 kΩ load)
12V at 20 mA ( < 600Ω load)
Zero and full scale adjustable from -99999 to +99999
16 bits (0.0015% of full scale)
250V rms working, 2.3 kV rms per 1 min test
(dual analog outputs share the same ground)

Relay Outputs (optional)

Relay Types

Current Ratings

Output common
Isolation

2 Form C contact relays or 4 Form A contact relays (NO)
2 or 4 Form A, AC/DC solid state relays (NO)
8A at 250 Vac or 24 Vdc for contact relays
120 mA at 140 Vac or 180 Vdc for solid state relays
Isolated commons for dual relays or each pair of quad relays
250V rms working, 2.3 kV rms per 1 min test