5

7750 series 4-20 ma transMiter

To convert the measured flow into a 4-20 mAdc signal, an angle of
rotation transmitter is mounted to the indicator. This device is fac-

tory calibrated to ensure accuracy and should only be adjusted by
King Instrument Company.

4-20 Ma transMitter speciFications:

POWER SUPPLY:

12-33V DC

MAXIMUM CURRENT CONSUMPTION: 40 mA
TEMPERATURE LIMITS:

-13°F to 158°F (-25°C to 70° C)

OUTPUT:

4 to 20 mAdc

ACCURACY:

<=0.5%

LINEARITY:

+/- 0.4%

INFLUENCE FROM BEARING:

+/- 0.1%

TEMPERATURE INFLUENCE (AMBIENT): +/- 0.03% per degree C
POWER SUPPLY INFLUENCE:

+/- 0.1%

LOAD RESISTANCE INFLUENCE:

+/- 0.03% at R max.

electrical connections:

7750 Series

Installation Instructions

12700 Pala Drive, Garden Grove, CA 92841

www.kinginstrumentco.com / (714) 891-0008

Meters are warranted against defects in materials and workmanship

to the original user for a period of thirteen (13) months from the date

of factory shipment, provided the meter is installed, operated and

maintained in accordance with King Instrument Company's

instructions and recommendations.

This warranty does not apply if failure is caused or contributed to by

any of the following: improper handling, improper storage, abuse,

unsuitable application of the product, lack of reasonable and

necessary maintenance, use exceeding suggested pressure and

temperature maximums, improper packaging for return, or repairs

made or attempted to be made by anyone other than King Instrument

Company, Inc.

KING INSTRUMENT COMPANY, INC. MAKES NO

WARRANTY AS TO THE FITNESS OF ITS PRODUCTS

FOR SPECIFIC APPLICATIONS.

This warranty is valid for the original end-user only and does not

apply to products that have been damaged or modified. This

warranty is non-transferrable and is limited to replacement or repair.

The liability of King Instrument Company arising out of its supply of

the products, or their use, shall not in any case exceed the cost of

correcting defects in the products as set forth above.

THIS WARRANTY IS A LIMITED WARRANTY AND

SHALL BE IN LIEU OF ANY OTHER WARRANTIES,

EXPRESSED OR IMPLIED, INCLUDING BUT NOT

LIMITED TO ANY IMPLIED WARRANTY OR

MERCHANTABILITY OR FITNESS FOR A PARTICULAR

PURPOSE. THERE ARE NO OTHER WARRANTIES

WHICH EXIST BEYOND THE DESCRIPTION OR FACE

HEREOF.

IN NO EVENT SHALL KING INSTRUMENT COMPANY BE

LIABLE FOR LOSS OF PROFITS, INDIRECT,

CONSEQUENTIAL OR INCIDENTAL DAMAGES.

Products should be returned, prepaid, to King Instrument Company,

Inc. with proof of purchase. Call factory for Return Merchandise

Authorization (RMA) number and return instructions.

2013A

To convert the measured flow into a 4-20 mAdc signal, an angle of rotation

transmitter is mounted to the indicator.

This device is factory calibrated to

ensure accuracy and should only be adjusted by King Instrument

Company.

4-20 mA TRANSMITTER SPECIFICATIONS:

POWER SUPPLY:
MAXIMUM CURRENT CONSUMPTION:

TEMPERATURE LIMITS:

OUTPUT:

12-33V DC

40 mA

-13°F to 158°F (-25°C to 70°C)

4 to 20 mAdc

ACCURACY:

<=0.5%

LINEARITY:

+/- 0.4%

INFLUENCE FROM BEARING:

+/- 0.1%

ELECTRICAL CONNECTIONS:

S1

SPAN

ZERO

-

+

OUT

V

TRANSMITTER HOUSING

TRANSMITTER SHAFT

TRANSMITTER

JUMPER WIRE

RED WIRE (+)

BLACK WIRE (-)

V-

V+

OUT

ext

R

Rext: EXTERNAL RESISTANCE =

POWER SUPPLY (V) -12V

OUTPUT SIGNAL (mA)

H

BROWN

BLUE

BLACK-1

RED-1

BLACK-2

(GROUND)

BROWN

BLUE

BLACK-1

RED-1

BLACK-3

BROWN

BLUE

RED-2

BLACK-2

ALARM

ALARM

SENSOR

ALARM

SENSOR

(HIGH)

(LOW)

All 7710 Series flowmeters may be fitted with one or two inductive slot sensors.

Inductive slot sensors are 2-wire, DC, low current devices and are designed to

be used with a remote barrier / switch isolator capable of powering the sensor

and providing the desired switching option(s). Barrier / switch isolators are

available with 220VAC, 110VAC or 24VDC supply voltage requirements, contain

single pole double throw (SPDT) relays, and are DIN rail mountable. (Only

24VDC units are actually powered by the rail.) See barrier / switch isolator

specifications for electrical connections and further details.

INDUCTIVE SLOT SENSOR-ELECTRICAL

SPECIFICATIONS

TYPE:

SUPPLY VOLTAGE:
OUTPUT:
OUTPUT LOAD CURRENT:

Inductive

5-25V DC (Switch Isolator)

NAMUR

<= 1mA-Float Present

>= 3mA (15mA Max.)-Float Absent

SWITCHING FREQUENCY:

2kHZ

HOUSING RATING:

IP67

WIRING:

2 Conductor, NAMUR

POS=BROWN / NEG=BLUE

TERMINALS: #1=POS / #2=NEG

PEPPERL+FUCHS

General Purpose

General Purpose

UL:

CSA:

SENSOR APPROVALS:

ELECTRICAL CONNECTIONS

RED-1

BLACK-1

(GROUND)

BLACK-2

(GROUND)

RED-2

BLACK-2

BLACK-3

(GROUND)

RED-1

BLACK-1

(GROUND)

SETTING ALARM TRIGGER POINTS:

1) Remove the eight screws securing the SHIELD.

2) Remove the SHIELD.

3) Carefully loosen the ALARM NUT. (Loosen just enough to rotate the ALARM

SENSOR BRACKET).

4) Carefully rotate the ALARM SENSOR BRACKET to the desired alarm set

point. If the meter has DUAL ALARMS, set the LOW ALARM set point and then

the HIGH ALARM set point.

5) Tighten the ALARM NUT.
6) Replace the SHIELD and secure with the eight screws.

TEMPERATURE INFLUENCE (AMBIENT):

+/- 0.03% per degree C

POWER SUPPLY INFLUENCE:

+/- 0.1%

LOAD RESISTANCE INFLUENCE:

+/- 0.03% at R max.

ZERO POINT MARK

ZERO POINT MARK

(2 WIRE ONLY)

TRANSMITTER - REAR VIEW

2-WIRE CONNECTION

ZERO: POTENTIOMETER P1 FOR

ZERO POINT

POTENTIOMETER P2 FOR

MEASURING RANGE AND

VALUE

SPAN:

SWITCH FOR REVERSING

ROTATION (NOT

APPLICABLE)

S1:

H: DC POWER SUPPLY (12V-33V)

SHIELD SCREW

SHIELD SCREW O-RING

SHIELD WITH GASKET

INDICATOR HOUSING/

POINTER HUB AND

ALARM NUT

ALARM SENSOR BRACKET

8X

8X

METER TUBE ASSEMBLY

POINTER NEEDLE

12700 Pala Drive, Garden Grove, CA 92841

www.kinginstrumentco.com / (714) 891-0008

Meters are warranted against defects in materials and workmanship

to the original user for a period of thirteen (13) months from the date

of factory shipment, provided the meter is installed, operated and

maintained in accordance with King Instrument Company's

instructions and recommendations.

This warranty does not apply if failure is caused or contributed to by

any of the following: improper handling, improper storage, abuse,

unsuitable application of the product, lack of reasonable and

necessary maintenance, use exceeding suggested pressure and

temperature maximums, improper packaging for return, or repairs

made or attempted to be made by anyone other than King Instrument

Company, Inc.

KING INSTRUMENT COMPANY, INC. MAKES NO

WARRANTY AS TO THE FITNESS OF ITS PRODUCTS

FOR SPECIFIC APPLICATIONS.

This warranty is valid for the original end-user only and does not

apply to products that have been damaged or modified. This

warranty is non-transferrable and is limited to replacement or repair.

The liability of King Instrument Company arising out of its supply of

the products, or their use, shall not in any case exceed the cost of

correcting defects in the products as set forth above.

THIS WARRANTY IS A LIMITED WARRANTY AND

SHALL BE IN LIEU OF ANY OTHER WARRANTIES,

EXPRESSED OR IMPLIED, INCLUDING BUT NOT

LIMITED TO ANY IMPLIED WARRANTY OR

MERCHANTABILITY OR FITNESS FOR A PARTICULAR

PURPOSE. THERE ARE NO OTHER WARRANTIES

WHICH EXIST BEYOND THE DESCRIPTION OR FACE

HEREOF.

IN NO EVENT SHALL KING INSTRUMENT COMPANY BE

LIABLE FOR LOSS OF PROFITS, INDIRECT,

CONSEQUENTIAL OR INCIDENTAL DAMAGES.

Products should be returned, prepaid, to King Instrument Company,

Inc. with proof of purchase. Call factory for Return Merchandise

Authorization (RMA) number and return instructions.

2013A

To convert the measured flow into a 4-20 mAdc signal, an angle of rotation

transmitter is mounted to the indicator.

This device is factory calibrated to

ensure accuracy and should only be adjusted by King Instrument

Company.

4-20 mA TRANSMITTER SPECIFICATIONS:

POWER SUPPLY:
MAXIMUM CURRENT CONSUMPTION:

TEMPERATURE LIMITS:

OUTPUT:

12-33V DC

40 mA

-13°F to 158°F (-25°C to 70°C)

4 to 20 mAdc

ACCURACY:

<=0.5%

LINEARITY:

+/- 0.4%

INFLUENCE FROM BEARING:

+/- 0.1%

ELECTRICAL CONNECTIONS:

S1

SPAN

ZERO

-

+

OUT

V

TRANSMITTER HOUSING

TRANSMITTER SHAFT

TRANSMITTER

JUMPER WIRE

RED WIRE (+)

BLACK WIRE (-)

V-

V+

OUT

ext

R

Rext: EXTERNAL RESISTANCE =

POWER SUPPLY (V) -12V

OUTPUT SIGNAL (mA)

H

BROWN

BLUE

BLACK-1

RED-1

BLACK-2

(GROUND)

BROWN

BLUE

BLACK-1

RED-1

BLACK-3

BROWN

BLUE

RED-2

BLACK-2

ALARM

ALARM

SENSOR

ALARM

SENSOR

(HIGH)

(LOW)

All 7710 Series flowmeters may be fitted with one or two inductive slot sensors.

Inductive slot sensors are 2-wire, DC, low current devices and are designed to

be used with a remote barrier / switch isolator capable of powering the sensor

and providing the desired switching option(s). Barrier / switch isolators are

available with 220VAC, 110VAC or 24VDC supply voltage requirements, contain

single pole double throw (SPDT) relays, and are DIN rail mountable. (Only

24VDC units are actually powered by the rail.) See barrier / switch isolator

specifications for electrical connections and further details.

INDUCTIVE SLOT SENSOR-ELECTRICAL

SPECIFICATIONS

TYPE:

SUPPLY VOLTAGE:
OUTPUT:
OUTPUT LOAD CURRENT:

Inductive

5-25V DC (Switch Isolator)

NAMUR

<= 1mA-Float Present

>= 3mA (15mA Max.)-Float Absent

SWITCHING FREQUENCY:

2kHZ

HOUSING RATING:

IP67

WIRING:

2 Conductor, NAMUR

POS=BROWN / NEG=BLUE

TERMINALS: #1=POS / #2=NEG

PEPPERL+FUCHS

General Purpose

General Purpose

UL:

CSA:

SENSOR APPROVALS:

ELECTRICAL CONNECTIONS

RED-1

BLACK-1

(GROUND)

BLACK-2

(GROUND)

RED-2

BLACK-2

BLACK-3

(GROUND)

RED-1

BLACK-1

(GROUND)

SETTING ALARM TRIGGER POINTS:

1) Remove the eight screws securing the SHIELD.

2) Remove the SHIELD.

3) Carefully loosen the ALARM NUT. (Loosen just enough to rotate the ALARM

SENSOR BRACKET).

4) Carefully rotate the ALARM SENSOR BRACKET to the desired alarm set

point. If the meter has DUAL ALARMS, set the LOW ALARM set point and then

the HIGH ALARM set point.

5) Tighten the ALARM NUT.
6) Replace the SHIELD and secure with the eight screws.

TEMPERATURE INFLUENCE (AMBIENT):

+/- 0.03% per degree C

POWER SUPPLY INFLUENCE:

+/- 0.1%

LOAD RESISTANCE INFLUENCE:

+/- 0.03% at R max.

ZERO POINT MARK

ZERO POINT MARK

(2 WIRE ONLY)

TRANSMITTER - REAR VIEW

2-WIRE CONNECTION

ZERO: POTENTIOMETER P1 FOR

ZERO POINT

POTENTIOMETER P2 FOR

MEASURING RANGE AND

VALUE

SPAN:

SWITCH FOR REVERSING

ROTATION (NOT

APPLICABLE)

S1:

H: DC POWER SUPPLY (12V-33V)

SHIELD SCREW

SHIELD SCREW O-RING

SHIELD WITH GASKET

INDICATOR HOUSING/

POINTER HUB AND

ALARM NUT

ALARM SENSOR BRACKET

8X

8X

METER TUBE ASSEMBLY

POINTER NEEDLE

2-wire connection

Rext: EXTERNAL RESISTANCE =

POWER SUPPLY (V) -12V

OUTPUT SIGNAL (mA)


H: DC POWER SUPPLY

(12V-33V)

transMitter - rear view

ZERO: POTENTIOMETER P1 FOR

ZERO POINT

SPAN: POTENTIOMETER P2 FOR

MEASURING RANGE AND

VALUE
S 1:

SWITCH FOR REVERSING

ROTATION (NOT APPLICABLE)