English 106

Error messages and display messages

Maintenance and service

Change the air filter (every year or if it is very dirty):

Turn the instrument off and disconnect the power cord.

Remove the fitted square air filter from the bottom of the instrument
and replace the dirty filter mat with a new one.

Clean the device

The outside of the instrument can be cleaned with a dry cloth or a mild
household cleaning agent. Under no circumstances may solvents such
as petroleum spirit, acetone or similar substances be used.
1.

Turn the instrument off and disconnect the power cord.

2.

When the instrument is cool, clean the surface of the instrument

with a soft, moist cloth and a weak soap solution. Make sure that

no water gets into the instrument.

Error message

Definition

Resolution

Prog.Mem. full

The user-specific program
memory is full

Delete a digestion program
that is no longer required

Prog.Mem. Clear

The user-specific program
memory is empty

Enter a user-specific
digestion program

Name not
recognised

No name has been entered

Designation

Name already
used esc

A name has been assigned
twice

Select a different name

Close the lid

The lid has not been closed

Close the lid

ERR: 1000
temp. too high

Permissible temperature
exceeded

Contact the Service
department

ERR: 1001
Catch

Lock faulty

Contact the Service
department

N O T I C E

Check the condition of your air filter (level of dirt) on a regular basis, as a very
dirty filter mat lengthens the cooling-down phase of the thermostat.

D A N G E R

Potential danger with contact with chemical/biological substances.
Working with chemical samples, standards and reagents can be dangerous.
Make yourself familiar with the necessary safety procedures and the correct
handling of the chemicals before the work and read and follow all relevant safety
data sheets.

WA R N I N G

Fire hazard! Do not use flammable agents to clean the instrument.
To prevent instrument damage, do not use cleaning agents such as turpentine,
acetone or similar products to clean the instrument or the display.

C A U T I O N

Burn hazard
Do not clean the instrument when it is hot.