Flashing 9s, Flashing 0s, Replacement parts and accessories – Hach-Lange 2100 AN IS Laboratory Turbidimeter User Manual User Manual
Page 47: Flashing 9s flashing 0s

To delete any calibration data entered by the user:
1. Turn off the instrument.
2. Push and hold CAL/Zero.
3. Turn on the instrument.
The CAL? light flashes. The instrument starts in Calibration mode.
4. Calibrate the instrument before use.
Flashing 9s
When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument. The
display will also flash 9s if Ratio is off and the measurement is greater
than 40 NTUs (1000 FNUs or 9.8 EBCs). Turn Ratio on. Refer to
Flashing 0s
When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument (i.e., less than
20 FAU) or a negative value. Calibrate the instrument.
• When measuring absorbance or transmittance, set the zero reference
point again.
• When measuring absorbance, make sure that the reading is positive.
To measure samples with negative absorbance, set the analytical
zero using the sample with the greatest absorbance and read the
sample with the least absorbance. Record the reading as negative
absorbance.
Replacement parts and accessories
Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Replacement parts
Description
Quantity
Item no.
Calibration kit, StablCal
®
, sealed sample
cells (<0.1, 20, 200, 1000, 4000 and
7500 NTU)
1
2659505
Cover, sample cell compartment
1
4702500
Cover, printer
1
4705400
Dust cover
1
4703000
Fuse for 115 V operation, time-delay,
250 V, 1.6 A, UL/CSA approved
1
3030700
Fuse for 230 V operation, time-delay,
250 V, 1.6 A, IEC type, VDE approved
1
3030600
Gelex
®
secondary turbidity standardization
kit
(stray light standard and 0–2, 0–20, 0–200,
200-4000, and 4,000–10,000 NTU)
1
2589200
Oiling cloth
1
4707600
Power cord, North America, 115 VAC,
UL/CSA approved
1
1801000
Power cord, European, 230 VAC, VDE
approved
1
4683600
Printer paper, thermal
5
4709000
Sample cells, 30mL, 1 in. round glass
6
2084900
Silicone oil
1
126936
English 45