Hach-Lange TSS PORTABLE Handheld Instrument Quick Guide User Manual
Quick guide, Tss portable handheld instrument, Turbidity measurement

1
DOC13.52.90141
Quick guide
TSS Portable Handheld Instrument
The instrument has two functions: turbidity measurement and suspended solids measurement. The appropriate
probe registered to the handheld instrument should be used for measurements. Refer to the safety information in
the manual before start-up.
An example of a turbidity measurement and a suspended solids measurement are given below. Refer to the
manual supplied for further information and details on the instrument.
Turbidity measurement
A standard curve C-TU is stored for the turbidity measurement. Calibration is not required.
1.
Connect probe to instrument.
2.
To switch the instrument on, press
ENTER/ON
for 2 seconds.
3.
Select
MEAS
and confirm with
ENTER/ON
.
4.
Place probe in measuring medium.
5.
Select
START
and confirm with
ENTER/ON
.
Note: The data is automatically saved every minute, max. 290 measurement points.
6.
To stop the measurement, select
MEAS
and confirm with
ENTER/ON
.
7.
Select
STOP
, confirm with
ENTER/ON
.
Delete the saved measurement points:
8.
Select
DATA
and confirm with
ENTER/ON
.
9.
Select
DELETE DATA
and confirm with
ENTER/ON
.
10. Select
YES
and confirm with
ENTER/ON
.
Suspended solids measurement
For the solids measurement, calibration is required. There are 4 curves available: C-DS1, C-DS2, C-DS3, C-DS4.
Each curve can be assigned to any measurement point. In the example, curve C-DS1 is selected.
1.
Connect the probe to the instrument.
2.
To switch on the instrument, press
ENTER/ON
for 2 seconds.
Select calibration curve
3.
Select
MENU
, confirm with
ENTER/ON
.
4.
Select
CALIBRATE
, confirm with
ENTER/ON
.
5.
Select
CURVE C-TU
and confirm with
ENTER/ON
.
Note: Curve C-TU is the default setting.
6.
Select calibration curve
C-DS1
with the navigation keys
UP/DOWN
and confirm with
ENTER/ON
.
The instrument reads the probe data and info text appears on the display.