Note: To print a diagnostic report, hold down PRINT, then turn the instrument on.

Table 4 Diagnostic codes

Code

Display

Description

21

Pr In

Printer test

22

Test results are shown.

Display test

23

Test results are shown.

Keyboard test

24

Test results are shown.

Memory test

Delete calibration data

To delete any calibration data entered by the user:

1. Turn off the instrument.
2. Push and hold CAL.
3. Turn on the instrument.

The CAL? light flashes. The instrument starts in Calibration mode.

4. Calibrate the instrument before use.

Flashing 9s

When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument.

Flashing 0s

When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument or a negative
value. Calibrate the instrument.