Hach-Lange 2100P iso - Portable Turbidimeter Instrument User Manual
Page 31

Section 5, continued
Page 31
4650098 diagnostics.fm
Error Messages
Message
Probable Cause
Corrective Action
E1
Dilution water is
≥ 0,5 FNU.
Start calibration over with better quality
dilution water or use a membrane filter to
filter the water before use.
E2
Two standards have the same value or
their difference is less than 60 FNU.
Not all standards were read during the
calibration.
Standard 1 is too low (<10 FNU).
Recheck preparation of standards and repeat
calibration.
E3
Low light error.
Re-read measurement.
Check lamp
Check for obstructed light path.
Dilution may be necessary.
E4
EEPROM malfunction.
Check sum failed. Press I/O. If E4 reappears,
call Hach service. If CAL? appears,
recalibrate.
E5
A/D overrange.
Check for obstructed light path.
E6
A/D underrange.
Check for open lid during reading and re-
read. Check for obstructed light path. If
persists, call Hach Service.
E7
Light Leak.
Close lid before pressing READ.
E8
Bad lamp circuit.
Reinsert lamp leads at terminal block-make
sure the lead ends are not touching each
other.If this fails, call Hach Service.
- AMTAX sc (118 pages)
- PHOSPHAX sc (106 pages)
- CL17 Instruction sheet (8 pages)
- CL17 USER MANUAL (60 pages)
- CLF10 sc (382 pages)
- 9184 sc (50 pages)
- POLYMETRON 9611 sc Installation (398 pages)
- POLYMETRON 9610 sc Maintenance and Troubleshooting (286 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SEQUENCER LINE INSTALLATION (2 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SS SAMPLE CONDITIONING (4 pages)
- HACH 5500 sc Installation (408 pages)
- HACH 5500 sc Maintenance and Troubleshooting (286 pages)
- HACH 5500 sc PO43-HR Operations (390 pages)
- HACH 5500 sc PO43-LR Operations (392 pages)
- HACH 5500 sc SiO2 Installation (418 pages)
- HACH 5500 sc SiO2 Operations (394 pages)
- POLYMETRON 9610 sc SiO2 Installation (410 pages)
- POLYMETRON 9240 Basic User Manual (162 pages)
- POLYMETRON 9240 Operator Manual (144 pages)
- PHOSPHAX sigma Basic User Manual (305 pages)
- AMTAX inter2 Basic User Manual (377 pages)
- GANIMEDE P (60 pages)
- BODTrak II Basic User Manual (172 pages)
- BODTRACK II User Manual (28 pages)
- BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) (46 pages)
- SPECTRO COLOR d-8 (29 pages)
- REFO 60_REFO 60D (36 pages)
- REFO 3_REFO 3D (19 pages)
- LICO 620 Basic User Manual (329 pages)
- LICO 620 User Manual (114 pages)
- SD900 (2 pages)
- LZX971 (8 pages)
- SC Sensor Plug and SC Sensor Coupling (8 pages)
- HQ Series Portable Meters (4 pages)
- LDO Sensor Replacement Kit for use with Hach HQ10 and HQ20 (43 pages)
- LDO Sensor Replacement Kit (8 pages)
- SIP 10 (273 pages)
- SC 100 (58 pages)
- SI792 E_T Quick start guide (2 pages)
- SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual (116 pages)
- SI792 D Quick start guide (2 pages)
- SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual (98 pages)
- SI794 C and SI794 D Short instructions (2 pages)
- SI794 D tr User Manual (108 pages)
- SI792 C Quick start guide (2 pages)