Basic user manual, Sensor specifications, Sensor product line – Hach-Lange ORBISPHERE 31xxx Basic User Manual User Manual
Page 3: Sensor membrane specifications, Hydrogen sensors, English

Sensor specifications
Specifications are subject to change without notice.
Sensor product line
Gas
Max. Pressure
rating (bar)
Sensor models Comments
O
2
20
31 11x.yz
Where :
x=Sensor special characteristics
(0 to 6 depending on application)
y=Membrane O-ring material
(0=EDPM; 1=Viton; 2=Kalrez; 4=Nitril)
z=Head material
(1=Stainless Steel; 2=Peek; 4=Hastelloy; 5=Titanium; 7= Monel)
Suffixes (when used):
A denotes a sensor with fast response to temperature change
E denotes an EEx certified sensor (Ex-Proof)
s denotes a Smart sensor
50
31 12x.yz
100
31 13x.yz
200
31 14x.yz
O
3
20
31 31x.yz
100
31 33x.yz
H
2
50
31 21x.yz
100
31 23x.yz
200
31 24x.yz
• All ORBISPHERE electrochemical sensor enclosures are certified IP68 / NEMA4
• PEEK (Polyetheretherketone) is a highly crystalline thermoplastic
Sensor membrane specifications
Hydrogen sensors
Specification
Membrane
2956A
Membrane
2952A
Membrane
2995A
Membrane
29015A
Recommended applications
Trace
measurement
Low
concentration
Average
concentration
High
concentration
Material
PFA
Tefzel
®
Tedlar
®
Saran
Thickness [μm]
25
25
12.5
23
Calibration gas
1% pure H
2
10% pure H
2
100% pure H
2
100% pure H
2
Dissolved measurement range
0 ppb to 75 ppb
0 ppb to
300 ppb
0 ppb to
3200 ppb
0 ppb to
32 ppm
Gaseous measurement range
0 Pa to 5 kPa
0 Pa to 20 kPa
0 Pa to 200 kPa
0 kPa to
2000 kPa
Accuracy
The greater of
±1% of reading or
± 0.03 ppb, or ±
1 Pa
The greater of
±1% of reading
or ± 0.09 ppb,
or ± 6 Pa
The greater of
±1% of reading or
± 1 ppb, or ±
50 Pa
The greater of
±1% of reading
or ± 10 ppb, or
± 1 kPa
Integrated radiation dose limit
2 x 10
4
10
8
10
8
N/A
Expected current in air @ 1 bar 25°C
[μA]
N/A
Expected current in pure gas [μA]
150
50
5
0.5
Temperature compensation range
0 to 50°C
0 to 50°C
10 to 45°C
10 to 45 °C
Temperature measuring range
-5 to 100° C
English 3