Section 6 troubleshooting, 1 error codes, 2 warnings – Hach-Lange SOLITAX sc User Manual
Page 21: 1 error codes 6.2 warnings

21
Section 6
Troubleshooting
6.1 Error Codes
In the case of an error, the indication of the measured value flashes on the display and all
the contacts and current outputs allocated to this sensor are placed on hold. The following
conditions will result in flashing measured values:
•
Data transmission between controller and sensor interrupted
On the Main menu open the SENSOR DIAG menu using ENTER and determine the
cause of the fault.
6.2 Warnings
A warning results in a flashing warning icon on the right of the display, all menus, contacts
and outputs remain unaffected and continue to work normally. On the Main menu open the
SENSOR DIAG menu using ENTER and determine the cause of the warning.
A warning may be used to trigger a relay and users can set warning levels to define the
severity of the warning.
Table 1 Error Messages
Error Displayed
Cause
Solution
POS. UNKNOWN
Wiper position unknown
Open the TEST/MAINT menu and trigger the ”WIPE” function, if the
problem persists contact the manufacturer's customer service
LED FAULTY
Faulty LED
Contact customer service
MOIST
Moisture value > 10
Remove the sensor immediately and store in a dry place, contact
customer service
CAL. DATA
Factory calibration data lost
Contact customer service
Table 2 Warnings
Warning Displayed
Cause
Solution
WARNING
Cause
Action
REPLACE WIPER
Counter elapsed
Replace wiper, reset counter
TEST/MAINT
Counter elapsed
Contact customer service
GASKET
Counter elapsed
Contact customer service