Section 1 specifications – Hach-Lange DR2800 eco User Manual
Page 5

5
Section 1
Specifications
Specifications are subject to change without notice.
Performance Specifications
Operating Mode
Transmittance (%), Absorbance and Concentration
Source Lamp
Gas-filled Tungsten (visible)
Wavelength Range
340–900 nm
Wavelength Accuracy
± 1.5 nm
Wavelength Reproducibility
< 0.1 nm
Wavelength Resolution
1 nm
Wavelength Calibration
Automatic
Wavelength Selection
Automatic, based on method selection
Spectral Bandwidth
< 8 nm
Photometric measuring range
± 3.0 Abs in Wavelength Range 340–900 nm
Photometric Accuracy
5 mAbs at 0.0–0.5 Abs
1% at 0.50–2.0 Abs
Photometric Linearity
< 0.5%–2 Abs
< = 1% at > 2 Abs
with neutral glass at 546 nm
Stray Light
< 0.2% T @ 340 nm with KV450/3
< 0.1% T @ 340 nm with NaNO
2
Data storage
200 measured values (result, date, time, sample ID, user ID)
Physical and Environmental Specifications
Width
220 mm (8.6 in)
Height
135 mm (5.3 in)
Depth
330 mm (12.9 in)
Weight
4.06 kg (8.95 Ibs) without battery
4.38 kg (9.66 Ibs) with battery
Operating Requirements
10–40 °C (50–104 °F), max. 80% relative humidity (non-condensing)
Storage Requirements
-40–60 °C (-40–140 °F) max. 80% relative humidity (non-condensing)
Additional technical data
Mains connection
15 V- / 30VA
Plug-in power supply unit: (100–240 V/50–60 Hz)
Interfaces
Use only shielded cable with max. length of 3 m.
1 x USB type A
1 x USB type B
Enclosure Rating
IP 41 with closed lid
IP 42 with Protective Cover in place
Safety class
Safety class II