Specifications

Specifications are subject to change without notice.

Specification

Details

Measurement method

VIS spectrophotometer for HACH and HACH LANGE water analysis
methods

Measurement mode

Transmittance (%), Absorbance (Abs) and Concentration (Conc)

Dimensions (W x D x H)

178 x 261 x 98 mm (7.0 x 10.3 x 3.8 in.)

Enclosure rating

IP67 (closed sample cell cover)

Weight

1.5 kg (3.3 Ib)

Power requirements (internal)

AA size Alkaline (4x) or
rechargeable Nickel Metal Hydride (NiMH) batteries (4x) (optional module
necessary

1

)

Power requirements (external)

Power supply: 110–240 VAC; 50/60 Hz (optional module necessary

1

)

Interface

USB mini (optional module necessary

1

)

Operating temperature

10 to 40 °C (50 to 104 °F); maximum 80% relative humidity (non-
condensing)

Storage temperature

–30 to 60 °C (–30 to140 °F); maximum 80% relative humidity (non-
condensing)

Lamp source

Xenon flash

Wavelength range

340–800 nm

Photometric measuring range

±3.0 Abs (wavelength range 340–800 nm)

Wavelength accuracy

±2 nm (wavelength range 340–800 nm)

Spectral bandwidth

5 nm

Photometric accuracy

3 mAbs at 0.0 to 0.5 Abs, 1% at 0.50 to 2.0 Abs

Photometric linearity

< 0.5% to 2 Abs
≤ 1% at > 2 Abs with neutral glass at 546 nm

Wavelength selection

Automatic, based on the method selection

Stray light

< 0.5% T at 340 nm with NaNO

2

Repeatability

± 0.1 nm

Wavelength resolution

1 nm

User programs (free programming) 50

Data logger

500 measured values (result, date, time, sample ID, user ID to GLP (Good
Laboratory Practice))

Sample cells

10 x 10 mm, 1 inch rectangular, 13 mm/16 mm/1 inch round, 1 cm/10 mL,
Flow-thru cell

Protection class

Power supply: Class II, instrument: Class III

Certifications

CE certified

Warranty

1 year (EU: 2 years)

1

Additional information is available on the manufacturer's website.