Specifications – Hach-Lange DR1900 User Manual User Manual
Page 5
Specifications
Specifications are subject to change without notice.
Specification
Details
Measurement method
VIS spectrophotometer for HACH and HACH LANGE water analysis
methods
Measurement mode
Transmittance (%), Absorbance (Abs) and Concentration (Conc)
Dimensions (W x D x H)
178 x 261 x 98 mm (7.0 x 10.3 x 3.8 in.)
Enclosure rating
IP67 (closed sample cell cover)
Weight
1.5 kg (3.3 Ib)
Power requirements (internal)
AA size Alkaline (4x) or
rechargeable Nickel Metal Hydride (NiMH) batteries (4x) (optional module
necessary
1
)
Power requirements (external)
Power supply: 110–240 VAC; 50/60 Hz (optional module necessary
1
)
Interface
USB mini (optional module necessary
1
)
Operating temperature
10 to 40 °C (50 to 104 °F); maximum 80% relative humidity (non-
condensing)
Storage temperature
–30 to 60 °C (–30 to140 °F); maximum 80% relative humidity (non-
condensing)
Lamp source
Xenon flash
Wavelength range
340–800 nm
Photometric measuring range
±3.0 Abs (wavelength range 340–800 nm)
Wavelength accuracy
±2 nm (wavelength range 340–800 nm)
Spectral bandwidth
5 nm
Photometric accuracy
3 mAbs at 0.0 to 0.5 Abs, 1% at 0.50 to 2.0 Abs
Photometric linearity
< 0.5% to 2 Abs
≤ 1% at > 2 Abs with neutral glass at 546 nm
Wavelength selection
Automatic, based on the method selection
Stray light
< 0.5% T at 340 nm with NaNO
2
Repeatability
± 0.1 nm
Wavelength resolution
1 nm
User programs (free programming) 50
Data logger
500 measured values (result, date, time, sample ID, user ID to GLP (Good
Laboratory Practice))
Sample cells
10 x 10 mm, 1 inch rectangular, 13 mm/16 mm/1 inch round, 1 cm/10 mL,
Flow-thru cell
Protection class
Power supply: Class II, instrument: Class III
Certifications
CE certified
Warranty
1 year (EU: 2 years)
1
Additional information is available on the manufacturer's website.
English 3