Flow-through accessories installation for NITRATAX sc
Sensor installation in the flow-through unit
CAUTION
Pinch hazard. Only qualified personnel should conduct the tasks described in this
section of the manual.
for the installation of the sensor in the flow-through unit.
1. Install the orientation screw to the sensor (
, item
, item
) to
make sure the measurement path in the measuring chamber is adjusted correctly.
Note: Replace the installed standard 50 mm sensor wiper with the supplied wiper with bore
wholes.
2. Insert the sensor (
) into the flow-through unit. Turn the sensor so
that the orientation screw slides into the cavity of the panel. Tighten the setscrew
(
1
, item
1
) to the sensor.
3. Connect the inflow and the outflow tubes (
). Insert the inflow tube
approximately 50 mm into the screw connection of the 2 and 5 mm sensors.
4. Use the allen key to seal the sensor to the measurement chamber.
Note: Remove the inflow tube before disassembling the 2 or 5 mm sensors.
Figure 2 Installation of flow-through unit for NITRATAX clear sc
1
Sample drain
4
Label (pay close attention to clearance requirements)
2
Ø 8 mm thru, clearance hole
5
Sample inlet
3
Spacer (2x) (wrap sensor cable)