Assembly
1. Attach a pair of color-code clips to the electrode and cable connector (
, item
2
).
2. Rinse probe with deionized water prior to use.
Calibration
For calibration steps, refer to the HQ Portable Meters User Manual. The IntelliCAL™ probes
store the current calibration and calibration history in the probe memory. It is not necessary to
recalibrate when moving a calibrated electrode from one meter to another.
Maintenance
General Maintenance
•
Between samples, wipe the probe with a clean towel.
•
If the connection to the meter is poor, readings may be unstable. To ensure accurate
readings, check the connection to the meter.
Storage
•
Between uses, store the probe dry in ambient conditions.
Specifications
Specifications are subject to change without notice.
Conductivity Range
0.01 µS/cm to 200.0 mS/cm
Conductivity Resolution
0.0–19.99 µS/cm: 0.01 µS/cm
20.0–199.9 µS/cm: 0.1 µS/cm
200.0–1999.0 µS/cm: 1.0 µS/cm
2.0–19.99 mS/cm: 0.01 mS/cm
20.0–200.0 mS/cm: 0.1 mS/cm
Conductivity Accuracy
± 0.5% of Reading
TDS Range
0 to 50,000 mg/L as NaCl
TDS Accuracy
± 0.5% of Reading
TDS Resolution
0.0–199.9 mg/L: 0.1 mg/L
200.0
–
1999.0 mg/L: 1.0 mg/L
2.0–19.99 g/L: 0.01 g/L
20.0–50.0 g/L: 0.1 g/L
Salinity Range
0 to 42 ppt (‰)
Salinity Accuracy
± 0.1 ppt (‰)
Salinity Resolution
0.01 ppt (‰)
Temperature Range
–10.0 to 110.0 ºC
Temperature Accuracy
± 0.3 ºC
Warranty
Sensor is covered by a one-year warranty.