Calibration – Hach-Lange LDO101-30 Instruction Sheet User Manual
Page 3

3
LDO101-05, LDO101-10, LDO101-15, or LDO101-30 Probe
Replacing the Shroud
1. Place the locking ring (
Figure 2
, item
2
) on the probe body with the threads toward the
shroud (
Figure 2
, item
1
).
2. Slide the shroud onto the probe until the locking ribs (
Figure 2
, item
4
) are seated in
the locking groove (
Figure 2
, item
7
).
3. Screw the locking ring onto the threads.
4. Hold the shroud and hand-tighten the locking ring.
Calibration
Each lot of LDO sensor caps is factory-calibrated. For best performance, a one-time
calibration initialization can be performed when a new sensor is installed. Additional
calibrations can be performed but are not required.
For calibration steps, refer to the HQ Portable Meters User Manual. The IntelliCAL™
probes store the current calibration and calibration history in the probe memory. It is not
necessary to recalibrate when moving a calibrated probe from one meter to another.
Figure 2 Shroud and Sensor Cap Removal
1
Shroud
6
Cap Seal
2
Locking Ring
7
Locking Groove
3
Sensor Cap
8
Temperature Sensor
4
Locking Ribs (4x)
9
Sensor Lens
5
O-ring