Maintenance, Clean the probe, Storage – Hach-Lange ISENO318103 User Manual User Manual
Page 10: Troubleshooting, If a measurement error occurs, refer to

6. Select Acceptance Criteria and update the settings:
Option
Description
Acceptance Limits
Sets the tolerance limits for check standard—1% to 20%.
Cal Expires on Failure Recalibration required if check standard fails—Yes or No.
The calibration expires if the check standard fails and Cal Expires is
set to Yes.
7. If prompted, enter a name for the new method settings. Additional changes made to
the settings of an existing method are automatically saved with the same method
name.
8. Push EXIT until the meter returns to the measurement mode.
Maintenance
Clean the probe
Clean the probe when:
•
Drifting/inaccurate readings occur as a result of contamination on the sensing
element or improper storage conditions.
•
Slow response time occurs as a result of contamination on the sensing element.
•
The slope is out of range as a result of contamination on the sensing element.
For general contaminants, complete the following steps.
1. Rinse the probe with deionized water. Blot dry with a lint-free cloth. Do not touch the
tip of the probe.
2. If harsh contaminants are attached to the probe, polish the probe tip with a soft cloth
or cotton swab to remove the contaminants.
3. Soak for 30 minutes in 100 mg/L Nitrate Nitrogen standard solution.
Storage
The probe can be stored dry. To protect the sensing element, rinse with DI water and blot
dry with a lint-free cloth. Install the sensor protection cap.
Note: The probe must be conditioned after long-term storage. Refer to
on page 2.
Troubleshooting
Message or symptom
Possible cause
Action
Probe not supported
Software not updated
To download the most current version of the
software, refer to the applicable product page on the
manufacturer's website.
Refer to the HQd Series meter manual for the
specific instructions for the meter model.
HQd meter does not support
IntelliCAL
®
probe
Contact a Technical Support Representative.