Hach-Lange 5337T User Manual
Page 5

Measuring range
2 to 14 pH
Operating temp.
0 to 80 °C
Max. pressure
2 bar
Reference element
cartridge
Diaphragm
open circular
Electrolyte
solid
Body material
glass
Thread
PG 13,5
Connector
MP-4
Temperature sensor
Pt 1000
Min. immersion depth
12 mm
5
GB
Storage
Place the electrode in the storage protec-
tor
with KCl + glycerine.
Life of an electrode
The average life expectancy of a pH proc-
ess electrode oscillates between 6 or 12
months, depending on the working con-
ditions.
Electrodes exposed to high temperatures
do not last as long. Very alkaline mediums
also reduces the life span of electrodes.
If an electrode has too short life span,
less than three months, ask us about it.
Probably there is a more suitable elec-
trode for the application.
Specifications
Applications
Dirty water, with mud, waste water, etc.
Limitations
Solutions with pH < 2, very clean water,
distilled and de-mineralised water.
Temperature > 80ºC. Pressure > 2 bar.
CERTIFICATE OF QUALITY
The attached electrode 53 37 T has
passed the quality test and follows the
next specifications:
• Asymmetry potential < ± 15 mV.
• Sensitivity, pH 4 to 7 (at 25 ˚C) > 98%.
• Response time, pH 4 to 7 < 20 s.
• Temperature error 0.5 ºC.