Product overview, Preparation for use – Hach-Lange PHC705 User Manual
Page 2

Specifications
Details
Storage solution
Saturated KCl
Minimum sample depth
18 mm (0.71 in.)
Dimensions
Body diameter: 12 mm (0.47 in.)
Body length: 103 mm (4.1 in.)
Total length: 200 mm (7.87 in.)
Cable length: 1 m (3.28 ft)
Cable connection
M12 digital output and connector compatible with HQd meters
1
Radiometer Analytical’s Red Rod technology: Encapsulation of reference elements with a red protection tube.
Product overview
The probe is a digital pH combination electrode with refillable electrolyte, a glass body
and an integrated temperature sensor.The probe is for laboratory use and measures pH
in alkaline samples (pH > 12).
Note: Do not use the probe with organic solvents, emulsions, or samples with high solids content.
A 59-mL bottle of reference electrolyte filling solution (saturated KCl solution) is included
with the probe. Refer to
.
Figure 1 Probe overview
1 Probe soaker bottle
5 Probe soaker bottle holder
2 Filling-hole cap
6 Reference junction
3 Protective tape and filling-hole
7 Glass bulb and temperature sensor
4 Cable, 1 meter (3.28 ft)
Preparation for use
Prepare the probe before a calibration or a sample measurement.
1. Turn the probe soaker bottle cap counter-clockwise to loosen the cap.
2. Remove the soaker bottle from the probe.
3. Fully rinse the reference junction and the glass bulb with deionized water to remove
all of the storage solution or possible salt particles. Dry with a no-lint cloth.
4. Remove the protective tape from the filling hole before initial use (refer to
on page 2). Discard the protective tape.
2