Hach-Lange HQD - Luminescent Dissolved Oxygen Probe LDO User Manual
User instructions, Safety information, Overview
1
DOC272.53.80021
User instructions
Luminescent Dissolved Oxygen Probe: Model LDO10101,
LDO10103, LDO10105, LDO10110, LDO10115 or LDO10130
Safety Information
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.
Overview
shows the LDO101 series probe, a luminescent dissolved oxygen (LDO
®
)
probe. The LDO10105, LDO10110, LDO10115 or LDO10130 rugged probe is available with a
5, 10, 15, or 30 meter cable. The LDO10101 or LDO10103 standard probe is available with a
1 or 3 meter cable. The probe provides measurement of the dissolved oxygen concentration in
wastewater, drinking water and general applications. For BOD applications, use the
LBOD10101 probe equipped with LDO technology and an incorporated stirrer.
Specifications
Specifications are subject to change without notice.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specification
Details
Dissolved oxygen range
0.1 to 20.0 mg/L (ppm) 1 to 200% saturation
Dissolved oxygen accuracy
± 0.1 mg/L for 0 to 8 mg/L
± 0.2 mg/L for greater than 8 mg/L
% saturation resolution
0.1%
Temperature range
0 to 50 °C (122 °F)
Storage temperature
0 to 40 °C (104 °F)
Response time
T90% at 10 sec (when stirred)
Temperature resolution
0.1 °C (32.18 °F)
Temperature accuracy
± 0.3 °C (32.54 °F)
Pressure resolution
1 hPa
Pressure accuracy
± 0.8%
Minimum sample depth
25 mm (0.984 in.)
Dimensions
Standard diameter: 15 mm (0.59 in.), Length: 103 mm (4.1 in.), Total length (standard):
200 mm (7.9 in.), Total length (rugged): 220 mm (8.7 in.), Cable length: 1 or 3 meter (3.3
or 9.8 ft.). Rugged diameter: 15 mm (0.6 in.), Length: 16 mm (0.63 in.), Cable length: 5,
10 15 or 30 meter (16.4, 32.8, 49.2, or 98.4 ft.)
Cable connection
Digital output and connector compatible with HQd meters
Warranty
Probe is covered by a 3 year warranty. Sensor cap is covered by a 1 year warranty.