Additional information, Specifications – Hach-Lange POLYMETRON 9586 User Manual
Page 5

Additional information
Additional information is available on the manufacturer's website.
Specifications
Specifications are subject to change without notice.
Panel
Specification
Details
Dimensions
817 x 300 x 140 mm (32.2 x 11.8 x 5.5 in)
Weight
14.6 kg (32.15 lbs)
Sample flow rate
10—15 liters/hour
Working pressure
0.5—6 bar (7—87 psi)
Ambient temperature
0—60 °C (-32—140 °F)
Relative humidity
10—90%
Working electrode
Platinum
Counter electrode
Stainless steel
Reference
Ag / AgCl / KCl 0.1 M
Measuring cell
Acrylic
Sample conditioner
DIPA (diisopropylamine)
European standards
EN 61326 Class A for EMC; EN 601010-1 for low voltage safety
International standards
cETLus
Sensor
Specification
Details
Measuring range
0—500 ppb hydrazine; 0—100 ppb carbohydrazide
Sensitivity
< 0.2 ppb
Repeatability
1 ppb or ± 2% of measurement whichever is the greater
Detection limit
≤ 1 ppb
Response time
< 60 seconds
Sample temperature measurement range
5—45 °C (41—113 °F)
Controller
Specification
Details
Component description
Microprocessor-controlled and menu-driven controller that operates the sensor and
displays measured values.
Operating temperature
-20 to 60 ºC (-4 to 140 ºF); 95% relative humidity, non-condensing with sensor load
<7 W; -20 to 50 ºC (-4 to 104 ºF) with sensor load <28 W
Storage temperature
-20 to 70 ºC (-4 to 158 ºF); 95% relative humidity, non-condensing
Enclosure
1
NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish
English 3