Hach-Lange LZU5399 User Manual
Page 5

Approx. constant
10 cm
-1
Measuring range
100 µS/cm to 500 mS/cm
Operating temp.
0 to 80 °C
Max. pressure
2 bar (25 ºC)
Temperature sensor
Pt 1000
Body material
glass
Electrode material
platinum
Thread
PG 13,5
Connector
MP-5
Min. immers. depth
50 mm
5
GB
Storage
Store the cell dry.
Life of the cell
The cell lifespan is unlimited if suitable
maintenance is applied and the cell does
not break.
Troubleshooting
... Reading 0 independently of the meas-
ured sample.
Broken cable.
Replace the cell.
... Unstable reading.
... Impossible calibration.
Dirty measuring electrodes or air
bubbles on their surface.
Clean or shake slightly the cell.
... The obtained readings differ a lot from
the real values.
Verify that the conductivity is covered
by the measuring interval of the cell.
Re-calibrate using standards in
good conditions.
.... slow cell response
Regenerate the cell.
Specifications
Applications
Samples with high conductivities, sea wa-
ters, industrial process waters, bleaches
and concentrated acids, etc.
QUALITY CERTIFICATE
The attached cell 53 99 has passed the
quality test and fulfils the following speci-
fications:
• EC error 5 %
between 100 µS/cm and 500 mS/cm.
• Temperature error 0.5 ºC.