Section 7 faults, causes, rectification, 1 error messages, 2 warnings – Hach-Lange 1200-S sc User Manual
Page 25: 1 error messages 7.2 warnings, Occurred (see section

23
Section 7
Faults, causes, rectification
7.1 Error messages
Possible sensor errors are displayed by the controller.
7.2 Warnings
Possible warning messages are displayed by the controller.
Table 1: Error messages
Error displayed
Cause
Rectification
*****
No communication with the
controller
Check the connection to the controller
Check the cable to the controller
Check the 12 V power supply
SENSOR MISSING
FFFFFFFFFFFFF
No communication with the
controller
Check the connection to the controller
Check the cable to the controller
Check the 12 V power supply
TEMP TOO LOW
Measured temperature < –5 °C
Ensure that the medium temperature is > –5 °C.
Check whether the internal resistance of the Pt 100 is
approx. 99 Ohm.
TEMP TOO HIGH
Measured temperature > +100 °C
Ensure that the medium temperature is < +100 °C.
Check whether the internal resistance of the Pt 100 is
approx. 138.5 Ohm.
pH TOO LOW
pH is lower than –2 pH
Check the sensor for leaks.
Change the electrode.
Change the board.
pH TOO HIGH
pH is higher than +14 pH
Check the sensor for leaks.
Change the electrode.
Change the board.
mV TOO LOW
mV is lower than –1500 mV
Check the sensor for leaks.
Change the electrode.
Change the board.
mV TOO HIGH
mV is higher than +1500 mV
Check the sensor for leaks.
Change the electrode.
Change the board.
Table 2: Warnings
Error displayed
Cause
Rectification
CAL TOO OLD
The last calibration was more than x days ago.
(Setting from the sensor setup)
Calibrate the sensor.
Set the calibration interval in the sensor setup.
HUMIDITY BAG
The desiccant bag is more than 1000 days old.
Please contact service.
REF IMP. LOW
Impedance of Reference electrode too low
Please contact service.
REF IMP. HGH
Impedance of Reference electrode too high
Please contact service.
GLASS IMP. LOW
Impedance of Glass electrode too low
Please contact service.
GLASS IMP. HGH
Impedance of Glass electrode too high
Please contact service.