General information – Hach-Lange SD900 User Manual User Manual
Page 6
Specification
Details
Storage/operating humidity
100% condensing
Pump
Peristaltic high speed, with spring-mounted Nylatron rollers
Pump enclosure
IP37
Pump tubing
9.5 mm ID x 15.9 OD mm (
3
/
8
in. ID x
5
/
8
in. OD) silicone
Pump tubing life
20,000 sample cycles with: 1 L (0.3 gal) sample volume, 1 rinse, 6 minute
pacing interval, 4.9 m (16 ft) of
3
/
8
in. intake tube, 4.6 m (15 ft) of vertical lift,
21 °C (70 °F) sample temperature
Vertical sample lift
Maximum 8.5 m (28 ft) for: 8.8 m (29 ft) of
3
/
8
-in. vinyl intake tube at sea level
at 20–25 °C (68–77 °F)
Pump flow rate
4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with
3
/
8
-in. intake tube typical
Sample volume
Programmable in 10-mL (0.34 oz) increments from 10 to 10,000 mL (3.38 oz
to 2.6 gal)
Sample volume repeatability
(typical)
±5% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m
(15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in. vinyl intake tube, single bottle, full
bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sample volume accuracy
(typical)
±10% of 200 mL sample volume using uncalibrated liquid detect with: 4.6 m
(15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in. vinyl intake tube, single bottle, full
bottle shut-off at room temperature and 1524 m (5000 ft) elevation
Sampling modes
Pacing: Time-fixed, flow-fixed, time-variable, flow-variable, constant time
variable volume (CTVV).
Refer to
Constant Time Variable Volume (CTVV) sampling
on page 23.
Distribution: Single bottle composite, multi-bottle composite, multi-bottle
discrete, bottles per sample, samples per bottle, combination of bottles per
sample, samples per bottle
Run modes
Continuous or non-continuous with user-entered number of samples
Multiple programs
Stores up to three sampling programs
Transfer velocity (typical)
0.9 m/s (2.9 ft/s) with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in. vinyl
intake tubing, 21 °C (70 °F) and 1524 m (5000 ft) elevation
Liquid sensor
Ultrasonic. Body: Ultem
®
NSF ANSI standard 51 approved, USP Class VI
compliant
Sample history
Up to 510 records
Air purge
Air purged automatically before and after each sample. The duration
automatically compensates for varying intake tube lengths.
Event log
510 records
Connections
Power, auxiliary, serial communications, distributor, SDI-12
Wetted materials
Stainless steel, polyethylene, Teflon, Ultem, silicone
Communications
RS232, Modbus, SDI-12
Warranty
1 year
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
4 English