b) Connect an ohm meter (100k ran ge) between "100k -1" and

"100k-2".

c) The meter should read 0 ohms.
d) Mo ve the slider gradually to th e right and observe the

ohmmeter.

e) The reading on the meter should increase from 0 to 100k ohms

i n proportion to the movement of the slide r.

f) Mo ve the ohmmeter lead from "100k -2" to "100k -3" an d check

for appro ximately 1 00k ohms.

8. BNC Connecto rs (BNC -1 & BNC -2)

(BNC-1 )

a) Check for continuity between "BNC -l-" and the metal shell of

BNC-1.

b) Check for continuity between "BNC -1+" and the center p in of

BNC-1.

c) Check for an open circuit between "BNC -1-" and "BNC -1 +".


(BNC-2 )

a) Check for continuity between "BNC -2-" and the metal she ll of

BNC-2.

b) Check for continuity between "BNC -2+" and the center p in of

BNC-2.

c) Check for an open circuit between "BNC -2-" and "BNC -2 +".



CHECKOUT COMPLETE