Specifications – Flowline AX18 Switch-Pro User Manual
Page 2

Step One
SPECIFICATIONS
2.8"
(71 mm)
3/4" NPT
(3/4" G)
2.3"
(57 mm)
2.1"
(54 mm)
2.8"
(71 mm)
2.8"
(71 mm)
7.5"
(191 mm)
Specifications:
Switch point:
1 (set by factory)
Orientation:
± 20˚ horizontal
Process temp.:
F: -40˚ to 194˚
C: -40˚ to 90˚
Pressure:
AU1_: 150 psi (10 bar)
AZ1_: 150 psi (10 bar)
Wetted material:
AU15: PP / PFA
AZ18: PP-Ryton®
Process mount:
3/4” NPT (3/4” G)
Enclosure rating:
NEMA 4X (IP65)
Encl. material:
PP, UL94VO
Conduit entrance: Single, 1/2" NPT
Termination:
2-4 poles
CE compliance:
EN 61326 EMC
EN 61010-1 safety
Sensor Technologies:
Ultrasonic (LU10 series)
Broadly applied in chemical, solvent, hydrocarbon and
light weight oil media
Supply voltage:
12-36 VDC
Consumption:
25 mA maximum
Contact type:
(1) SPST relay
Contact rating:
GP: 120 VAC/VDC @ 1A
IS: 32 VDC @ 0.5A
Contact output:
Selectable NO/NC
Classification:
Intrinsically safe
Vibration (LZ12 series)
Typically applied in wastewater media with light coat-
ing and/or foaming characteristics
Supply voltage:
12-36 VDC
Consumption:
25 mA maximum
Contact type:
(1) SPST relay
Contact rating:
120 VAC/VDC @ 1A
Contact output:
Selectable NO/NC
Classification:
General purpose
2
.8
"
(7
1
m
m
)
3/4" NPT
(3/4" G)
2.3"
(57 mm)
2.3"
(57 mm)
2.8"
(71 mm)
3.0"
(76 mm)
7.7"
(196 mm)
Dimensions (w/ ultrasonic sensor):
Dimensions (w/ vibration sensor):