SPECIFICATIONS
MODEL 1655A
OUTPUT ISOLATION
Leakage less than 0.1mA (25ºC, 50% relative humidity.)
VOLTAGE ADJUSTMENT RANGE
0-150VAC, with input at 120VAC
VOLTAGE/CURRENT SENSING
Sine wave average, calibrated in RMS
MAXIMUM CURRENT (Isolated)
3A continuous, 4A intermittent (0-130V). Intermittent not to
exceed 4 minutes at 4A followed by 5 minutes at 1A max.
METER SCALES
Voltage ....................................................................:0-150V
Current:. .......................................................................0-2A
.........................................................................0-4A
Volt-Amps (Voltage Set to 120)............................ 0-240VA
................................................................. 0-480VA
Leakage ......................0 to 5mA (5,000µA), expanded scale
METER ACCURACY (25ºC)
Volts:.........±5%
(typical) of full scale (calibrated @ 120VAC)
Current:.............................................±5%
(typical) of full scale
Leakage:............................................................±5% @ 500µA
MODEL 1653A
OUTPUT ISOLATION
Leakage less than 0.1mA (25ºC, 50% relative humidity)
VOLTAGE ADJUSTMENT RANGE
0-150VAC, with input at 120VAC
VOLTAGE/CURRENT SENSING
Sine wave average, calibrated in RMS.
MAXIMUM CURRENT (Isolated)
2A continuous (0-130V).
METER SCALES
Voltage:.........................................................................0-150V
Current:.............................................................................0-2A
METER ACCURACY (25ºC)
Volts:........±5%
(typical) of full scale (calibrated @ 120VAC)
Current:............................................±5%
(typical) of full scale