Model 2100
1: Mount the pressure transmitter onto a 1
in pipe (1” NPT thread) and mount the
pipe firmly at the desired measuring
location.
2: Lower the pressure transmitter into the
liquid. Take care not to hit the bottom
hard since it may damage the transmit-
ter!
1: Mount a suitable hook over the desired
measuring location. Note the weight of
the cable.
2: Remove the inner conical sleeve from
the cable fitting and pull the cable
through the outer part. Open the inner
sleeve and fit it around the cable at the
desired fixation point and press the
inner sleeve into place in the outer part.
Secure the cable fitting by pulling the
cable downwards.
3: Lower the pressure transmitter into the
wellpipe. Take care not to hit the
bottom hard since it may damage the
transmitter!
Mechanical mounting,
Model 1100
Model 3100
1: Mount a suitable hook over the desired
measuring location. Note the weight of
the cable.
2: Mount the cable fitting onto the cable.
Open the fitting by sliding the two plastic
jaws upwards, place the cable between
the jaws and slide the jaws downwards
until the cable locks. Secure the cable
fitting by pulling the cable downwards.
3: Lower the pressure transmitter into the liquid.
Take care not to hit the bottom hard since it
may damage the transmitter!
4: If the transmitter is to be used in a
wetwell or other locations with
turbulence or other disturbance, it is
advisable to install a pipe (min. nominal
diameter = 3 in) to protect the trans-
mitter from bumping into the wall or
other components. It is very important
that minimum 1 in of the pressure
transmitter is not being covered by
the pipe!
Cable fitting
Protective pipe
Model 2100
Model 1100
Cable fitting
Min. nominal
dimension = 3 in.
Min. 1 - 2 in !
1 in NPT thread
Model 3100