Maintenance and cleaning
SenTix
18
ba75843e01
08/2009
Cleaning
Remove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:
Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.
After cleaning
Rinse the electrode with deionized water and condition it in reference
electrolyte solution for at least 1 hour. Then recalibrate the electrode.
Contamination
Cleaning procedure
Fat and oil
Rinse with water containing household wash-
ing-up liquid
Lime and hydroxide deposits Rinse with citric acid (10 % by weight)
Proteins
Immerse in pepsin cleaning solution PEP/pH
for approx. 1 hour.
Note: Make sure the level of the reference
electrolyte is above that of the cleaning solu-
tion.